| Hioki Recorders
|
 |
8807-51 MEMORY HiCORDER
|
| Overview |
| Instantaneous Analysis and Long-term Recording of Harmonic Waves for Maintenance of Commercial Power Systems
|
- Measure harmonics up to 40 orders from the fundamental wave
- Analysis display includes RMS value, content factor, phase angle, active power, and power phase angle for each order of harmonics (numeric and graphic display)
- Analysis display of total RMS value, total deitortion, active/ reactive/ apparent power, and power factor (numeric display)
- Bar graph and numeric data display
- Power phase angle can be displayed as a vector
|
|
 |
 |
 STANDARD |
 Commercially available |
|
| Options |
MEMORY HiCORDER & MEMORY HiLOGGER Application Disk
- Description of transfer commands used with the GP-IB, RS-232C and LAN interfaces with WaveViewer
|
Check out
these other options for parts
8807-51
|
|
| Basic specifications
|
| Measurement ranges |
10 mV to 100 V/division, 13 ranges (10 division f.s.), resolution: 1/160 of range, AC voltage for possible measurement/ display using the memory function : 450 V rms
100 V/division excludes the RMS & memory recorder function
|
| Frequency characteristic |
DC to 50 kHz ±3 dB |
Time axis at memory recorder function
|
200 µs to 5 minutes/division, 19 settings (1 division =80 samples)
|
| Measurement functions |
(1)Memory recorder, (2)Recorder, (3)RMS recorder & memory (50/60 Hz or DC only)
(4)Harmonic wave analysis (Time series analysis mode), (5)Harmonic wave analysis (Instantaneous analysis mode)
|
| Number of input channels |
Analog (2 channels standard) + logic (8 channels standard)
*fixed input section |
| Memory capacity |
256 k words/ch (using CH1) to 128 k words/ch (using CH1 & CH2)
|
| Data storage |
Type II PC card slot (9726/ 9727/ 9728/ 9729 PC card, 128MB to 1GB)
|
| Interfaces |
RS-232C (the optional 9612 RS-232C CABLE is required for connection to PC) PC card: Commercially available PC card type fax modem (Please contact HIOKI for information on compatible fax modems)
|
Interfaces (optional, sold separately)
|
Detachable printer: 8992 PRINTER UNIT can be connected
|
| Recording and display |
5.7-inch STN color LCD, 112 mm (4.4 in) × 18 m (59.06 ft), thermal paper roll
|
| Other functions |
DMM function (voltage shown as numerals on the display), Scaling, Parameter calculations, Cursor readout, etc.
|
| Power supply |
100 to 240V AC (50/60 Hz) with use the 9418-10 AC ADAPTER (option)
9447 BATTERY PACK (Continuous use 3 hours) (option)
LR6 (AA) × 6 (Continuous use 1 hours) (LR6 batteries cannot be used with 8992 PRINTER UNIT)
|
| Dimensions, mass |
203 mm(7.99 in)W × 170 mm(6.69 in)H × 52 mm(2.05 in)D, 1.1 kg (38.8 oz)
280 mm(11.02 in)W × 170 mm(6.69 in)H × 52 mm(2.05 in)D, 1.5 kg (52.91 oz) (printer attached)
|
| Accessories |
LR6/AA alkaline batteries (6), alkaline battery box (1), shoulder belt (1), Wave viewer software (1)
|
|
| Specifications for the harmonic wave analysis section
|
| *The waveform recording functions of the 8807-51 is identical to these of the 8807-01. For details on specifications other than these concerning the harmonic wave analysis functions, refer to the catalog for the 8807-01
|
| Connection types |
Independent channels, single-phase two-wire
|
| Fundamental frequency range
|
45 Hz to 65 Hz, Automatic setting or manual setting (0.1 Hz resolution)
|
| Number of orders for analysis
|
Fundamental wave to 40th order |
| Analysis frequency band |
45 Hz to 2.6 kHz |
| Sampling speed |
400 kS/s fixed |
| Number of FFT operations |
512 points (sampled during one cycle of the fundamental wave)
|
| Time series analysis mode |
For any harmonic wave order number:
RMS value/ content factor/ phase angle/ active power/ active power content factor/ power phase angle
Total RMS value, total distortion-F, total distortion-R, active power, apparent power, reactive power, power factor
|
Time axis at time series analysis mode
|
5 minutes to 12 hours/division, 7 settings (80 samples/division)
|
| Instantaneous analysis mode
|
Numeric data + graph: harmonic wave RMS value, harmonic wave content factor, harmonic wave phase angle, harmonic wave active power, harmonic wave active power content factor, harmonic wave power phase angle
Numeric data only: Total RMS value, total distortion-F, total distortion-R, active power, apparent power, reactive power, power factor
|
| Other functions |
Harmonic wave trigger function, scaling, wiring and level check, over-range function, tabulated results output, cursor measurement
|
|
|
|