connex-electronics.com Hioki
Hioki
Hioki Environmental Measuring Instruments
Add part number 3418 to your Quote. 3418 TEMPERATURE HiTESTER
Overview
Quick and easy, a hand-held non-contact temperature measurement
These models can measure surface temperature of an object without coming into contact with it. This is done by sensing the infrared radiation energy which all bodies emit in accordance with their temperature.
  • Without laser marker
  • -50 °C to 500 °C
  • (1 °C resolution)
  • Ø72 mm at a 1 m distance
CE
Options (Click for picture)
Used for accurate measurement of the temperature of objects which have low thermal emissivity(e) such as polished metal, and for calibrating thermal emissivity
BLACK BODY SPRAY - e=0.95, 180 ml, 2-coat coverage about 1.0 m 2 , good to 550 °C BLACK BODY TAPE - e=0.95, 50 mm x 10 m, 1 roll, good to 180 °C
Basic specifications
Measurement range -50 °C to 500 °C, 1 °C resolution
Measurement field diameter (normal) Ø 72 mm(2.83 in) at a distance of 1 m(3.28 ft)
Accuracy ±1 % rdg. (at 201 to 500 °C), ±2 °C (at 0 to 200 °C), ±10 % rdg. ±2 °C (at -50 to -1 °C)
Response time 1.5 seconds (95 % response)
Other functions Upper and lower alarm, Maximum /minimum display, Auto power OFF, etc.,
Power supply R03(AAA) × 4 (Continuous use of 40 hours)
Dimensions, mass 40 mm(1.57 in)W × 170 mm(6.69 in)H × 36 mm(1.42 in)D, 140 g (4.9 oz)
Accessories Carrying case (1)
Back