| Basic specifications
|
| Measurement mode |
Earth leakage current Leakage current between enclosure and earth /
Leakage current between enclosure and enclosure /
Leakage current between enclosure and line / Patient leakage current I / Patient leakage current II /
Patient leakage current III / Patient auxiliary current
|
| Target current |
DC / AC / AC+DC (25mA max), AC peak (75mA max)
|
| Measurement range |
DC / AC / AC+DC mode ; 50 µA / 500 µA / 5 mA / 25 mA
AC peak mode ; 500 µA / 1 mA / 10 mA / 75 mA
|
| Measurement system |
Indication of a current value calculated based the measured drop in voltage caused by simulated resistance of the human body.
Measurement of true effective value. The measurement section ; chassis-grounded and floating.
|
| Accuracy |
DC / AC / AC+DC mode ; ± (2.0 % rdg. + 6 dgt.)
AC peak mode ; ± (2.0 % rdg. + 2 dgt.) DC mode ; ± (2.0 % rdg. + 3 dgt.)
|
| Input resistance |
1 M-ohm ± 1 % (Excluding voltmeter section, simulated resistance of the human body)
|
Network (human simulated resistance)
|
For medical electrical equipment / For IEC 60990 /
For JIS / For UL / General-purpose 1 / General-purpose 2 |
| Functions |
110 % voltage application function / Wiring check function / Automatic measurement function / Application line selection function / Ground fault prevention function / Setting of single-fault condition / Switching power supply polarity / Setting of measuring
time / Measurement delay / Maximum value hold / Allowable value judgement / Data save / Clock / Data back up / etc.
|
| Interface |
EXT I/O , RS-232C , GP-IB |
| Power supply |
100, 120, 220, 240 V AC (default setting) (50/60 Hz, Rated power ; 30VA)
|
| Dimensions, mass |
320 mm(12.6 in)W × 110 mm(4.33)H × 263 mm(10.35)D, 4.0 kg(141 oz)
|
| Accessories |
9170 TEST LEAD(2), 9195 ENCLOSURE PROBE(1), 9399 CARRYING CASE(1), Alligator clip(3)(2 red, 1 black), AC Power cord(2), Spare fuse(2)
|